8. Operating Manual
This operating manual is a work in progress.
IF YOU FEEL UNCERTAIN ABOUT
PERFORMING ANY OF THE PROCEDURES DESCRIBED BELOW -- SEEK HELP.
Topics
8.1. Introduction
8.2. Optical Microscope
8.3. Probe Start-up
8.4. Sample Change
8.5. Performing Analyses
8.6. Storing Line
Traverses
8.7. Acquiring BSE Images
8.8. Using the EDS System
8.9. Acquiring X-ray Dot Maps
8.10. Probe Shutdown
8.1. Introduction
The following conventions are used throughout this write up:
(1) labels on the scanning console are given in bold; (2) items entered
on or done with the Mac IIvx that serves as a terminal for the MICRO computer
are shown in underlined; and (3) items entered on the MICRO keyboard are
given in square brackets [ ].
8.2. Optical Microscope
The sample may be viewed in either reflected or transmitted
light, which are selected using the toggle switch to the right of the oculars.
The middle position of the switch is off. There is only one magnification
available: 400x. Each tick mark on the reticule corresponds to 1 µm.
8.3. Probe Start-up
- Fill the liquid N2
dewar. If the day is either Monday or Friday, fill the dewar and
record this fact in the log book.
- Turn on the Mac G3 and MAC
IIvx. Press the arrow key in the upper right corner of each
keyboard.
- Check the spectrometer and
stage positions. The spectrometers should be located at
approximately 600 and the stage at X=100, Y=100. If the spectrometer
positions are incorrect, seek help. If the stage X and Y positions are
incorrect, they can be adjusted by disengaging the motors by pulling out
the clutch knobs and moved. Be sure to reengage the clutches when you are
finished. Press [INIT] twice on the MICRO keyboard.
- Check the vacuum is adequate.
The VACUUM CONTROL second. vac.
light should be on. If not, check whether the gate value is open, If the
gate value is closed, open it using the switch labeled VANNE SECONDAIRE in the back of the
scanning console. The gate valve's closed position is marked. The gate
valve, which is located behind the electron column, will open (it is a
noisy process). Wait until the VACUUM
CONTROL second. vac. light comes on, before proceeding.

- Turn on the high voltage.
Turn the VOLTAGE coarse knob.
Each click will be accompanied by a grinding noise as the bias resistors
are changed. Position 5 corresponds to 15 keV.

- Start the control program.
Double-click the MBX 1.41b alias icon.
- Move the stage to a
fluorescent material. Benitoite (BENI), scheelite (SCHE), or
willemite (WILL) are good choices. Click: Standard (Stage Control
window), enter the standard abbreviation, then click Move. After the
stage is done moving, check the focus.
- Saturate the filament.
Do not exceed the previous saturation value recorded in the logbook.
- Turn the range on the ABSORBED CURRENT display to
3 x 10-8.
- Depress the button below the filament temperature
knob.
- Increase the filament temperature potentiometer
gradually while watching the ABSORBED
CURRENT meter. The absorbed current will increase and then
stop increasing as it reaches a plateau. The filament is
"saturated" at the leading edge of this plateau. If the
specimen current drops off after reaching a peak, the electron gun needs
aligning. Use the horizontal knobs on the side of the gun housing to
maximize the current, then resaturate the filament.
- Check the column alignment.
- Look at the fluorescent spot in the optical microscope.
- Turn the LENS 3
coarse potentiometer both ways to increase and decrease the size of
the spot. The fluorescent area should expand and contract symmetrically
from a point when you adjust the potentiometer up and down.
- If it does not, the column needs aligning. Adjust the
position of the whole column by using the black knobs halfway up the
electron column. Recheck the fluorescent area while adjusting the
LENS 3 coarse potentiometer. Repeat
the process until the expansion/contraction is symmetrical.
- Align the optical microscope.
Adjust the two slanting knobs at the base of the electron column so that
the spot is under the crosshairs when viewed using the eyepieces.
- Verify the spectrometer
positions. Move to andradite (ANDR). Focus. Adjust the absorbed
current to about 25 nA using the LENS 1
coarse potentiometer. Click: Functions > Spectrometer Controls
> Verify to perform the verification. Repeat until all values are
between 3 and negative 3.
8.4. Sample Change
- Turn off the filament.
Turn the filament temperature potentiometer to zero.
- Switch to the airlock vacuum
gauge. Turn the VACUUM GAUGES
knob to PRIMARY GAUGES: air lock.
- Attach the sample changer.
Remove the airlock cover plate and attach the sample change unit by
hanging it from the two pins at the top.
- Pump down the sample changer.
Press the AIRLOCK pump down
button on the VACUUM CONTROL
panel. The changer will pump out. Proceed when the vacuum is less than 2
Torr.
- Open the sample chamber door.
Hold the sample changer knob back while opening the sample chamber door.
Turn the handle forward, up, then left.
- Insert/remove the sample.
Gently push in the sample changer rod until it seats. Rotate and withdraw
it, levaing/capturing the sample. Be sure to pull the rod all the way out.
- Close the sample chamber door.
Hold the sample changer knob back while closing the sample chamber door.
Turn the handle forward, down, then right.
- Vent the sample changer.
Press the AIRLOCK air inlet
button on the VACUUM CONTROL
panel. The changer will vent and can be removed.
- Remove the sample changer.
Remove the sample changer and replace the airlock cover plate.
- Turn off the airlock system.
Press the AIRLOCK off button
on the VACUUM CONTROL panel.
- Switch to the secondary vacuum
gauge. Turn the VACUUM GAUGES
knob to SECONDARY GAUGE.
-
Choose your analytical setup.
Enter Apple-L and select the desired procedure from the "Analytical
Procedures" subdirectory.
-
Search the Standard Intensities
Library. Choose the option that does a search for any more recent
records (calibrations) for the the current setup. Click OK.
-
Set the beam current for analysis.
Use the LENS 1 coarse
potentiometer to set the beam current to the desired value. For most analyses
this will be 10 nA.
-
Specify
output file. Enter an
output file name. It's a good idea to identify the sample in the file name.
-
Record run information in the log
book. Write your name, the date, the time, the KV used, the filament
temperature value, the beam current, the spot size, and the column vacuum.
-
Check the standards. Check
that the K-ratios for the standards are within 1% of 1.000. If not,
recalibrate:
-
Enter Apple R and select the element(s) and standard(s) to
recalibrate. The MICRO will move to the appropriate standard automatically.
-
Position and focus the standard carefully, then hit [ ADJU ].
-
You must repeat this three times, after which, the computer will
report the results.
-
Accept the calibration if the Sigma Ratio is less than 2.0.
-
Check secondary standard(s).
Move to a standard similar to your unknowns (if possible) and perform an
analysis to check the calibration. If any major elements are off by more than
2% (relative), you may wish to recalibrate the standard (see above).
-
Perform analyses. Click the
"Use previous backgrounds" box after performing the first analysis on
an unknown. You may wish to unclick this box the first time you go to a new
material; however, backgrounds should be pretty similar for all materials if
selected correctly.