4.4. Accuracy and Precision
Accuracy may be defined as how close a measured value is
to the "true" value. It is often difficult to establish a "true"
value and thus the accuracy of an analysis based upon it. Accuracy is affected
if the compositions of the standards are not well known whereas precision is
not. Standard compositions are often determined by wet chemistry, but this
process also analyzes tiny inclusions in the standard materials making it
impossible to establish "true" composition.
Sometimes for the minor
elements, microprobe analysis of inclusion free areas in a mineral is the best
method for determining the true composition! One way to avoid potential
problems with the standard compositions is to use mono-elemental (synthetic)
standards. With these materials, one only has to look for impurities and make
sure that they are insignificant, not determine their abundances. However, the
matrix effects may be so significant that a multi-element standard (similar to
the unknown) is required.
Precision refers to how well a given
measurement or results can be reproduced. Values can be very precisely
determined and still be very inaccurate. Conversely, a number of imprecise
analyses may average to a very accurate value. Precision is effectively limited
by counting statistics when dealing with X-ray analysis.
Many factors, many out of the control of the
analyst, can affect both precision and accuracy. Among them are:
- Incorrect standard values,
which affect accuracy and produce systematic errors (high values if
standard compositions are higher than they should be, lower if low).
- Focus problems, which can
produce significant unsystematic errors and a loss of precision and
accuracy. Defocusing the beam on the sample results in imperfect
spectrometer optics and reduced count rates. This produces unsystematic
errors unless one always misfocuses identically.
- Specimen tilt, which
produces systematic errors by changing the take-off angle.
- Irregularities in the
sample surface, which are produced during polishing, or despite polishing,
also may produce unsystematic errors.
- Errors in matrix-correction
factors, which can significantly reduce accuracy especially where
correction coefficients are poorly known. The element F is a good example
of this--F in apatite should be determined using a fluorapatite standard,
while determining F in micas requires a F-rich mica. Errors in the
matrix-correction coefficients will produce systematic errors during data
reduction. Everyone worries about getting better (correct) data-reduction
factors, but not many do anything about it.
- Variations in the thickness
of the C-coat, which primarily affect accuracy and cause variable amounts
of absorption for soft X-rays. The largest concern for routine analysis is
the effect of coat thickness on Na-Ka X-rays. This
problem can be significant if standards and unknowns have very different
thicknesses of carbon, and will cause systematic variations.
- Incorrect location of
backgrounds, which can strongly affect accuracy as can large deadtimes and
peak shifts from standard to unknowns (e.g., S).
- Errors in nominal
accelerating voltage, which cause systematic errors. The gun accelerating
voltage can be checked using the excitation of certain X-ray lines (e.g.
Ti-Ka at 4.964 keV and Rb-Ka at 15.200 keV).
- Electronic instability,
which primarily reduces precision.
Copyright 1997-2003,
James H. Wittke
Last update:
01/18/2006 01:47 PM.